UltraRAE 3000 (Model PGM-7360) Photoionisation (PID) Detector

Dual-detection detector for benzene or over 200 VOCs.  Use RAE-SEP benzene tubes when the unit is in benzene mode to gain highly specific readings.  Suitable for applications including entry pre-screening during refinery and plant maintenance to hazardous material response, marine spill, response and refinery down-stream monitoring.

Manufacturer:

UltraRAE 3000+ (PGM-7360) Photoionisation (PID) Detector with 9.8 eV lamp Non-WirelessThe UltraRAE 3000 from RAE Systems is a cutting-edge handheld VOC monitor designed to provide accurate and reliable readings in critical applications. Whether you’re ensuring safety in confined spaces, conducting refinery and plant maintenance, responding to hazardous material spills, or performing marine and plant overhauls, the UltraRAE 3000 delivers the precision and functionality you need.
The UltraRAE 3000+ from RAE Systems is a photoionisation (PID) detector, suitable for the detection of hazardous and volatile organic compounds (VOC’s) The PID is capable of detecting more than 200 VOCs, but also features a benzene-specific mode and is Non-Wireless.
This makes the UltraRAE 3000 ideal for pre-screening applications in refinery and plant maintenance, hazardous material response, marine spill response and refinery downstream monitoring. The detector has an extended range of 0.05 to 10,000ppm in VOC mode and 50 ppb to 200ppm in its benzene-specific mode.
The UltraRAE 3000 comes with a 9.8 eV lamp as standard, but other lamps are available. The sensor responds extremely quickly and can produce highly specific readings when using RAE-SEP benzene tubes.
Save money, buy the UltraRAE 3000+ (PGM-7360) Photoionisation (PID) Detector with 9.8 eV lamp with the accessories kit
The UltraRAE 3000 combines state-of-the-art technology, ease of use, and low ownership costs, making it the ideal choice for your VOC monitoring needs. Available from Frontline Safety with additional accessories like RAE-Sep benzene tubes—your one-stop solution for reliable gas detection. 

Key Features

Easy to read, graphical display
Multi-language support
Maintainable with easy access to sensors, lamp and battery without tools
Specification method reduces risk of false alarms
Alarm system – flashing light, 95dB buzzer
IP 65 rated (unit running)
Datalogging – 6 months at one-minute interval

What’s Included?

• UltraRAE 3000+ Monitor, Model PGM-7360 with soft leather carrying case
• Datalogging with ProRAE Studio II Package for Windows ® XP, Windows® Vista, or Windows® 7
• Travel Charger
• RAE UV lamp and RAE-Sep Tubes
• Flex-I-Probe
• External filter
• Red rubber boot
• Alkaline battery adapter
• Lamp-cleaning kit
• Tool kit
• Lithium-ion (Li-Ion) battery with universal AC/DC charger and international plug kit
• Operation CD-ROM
• Operation & Maintenance manual

 

Sensor

Photoionization sensor with standard 9.8 eV or 11.7 eV lamp

Range

VOCs – 0 to 99.99 ppm
              100 to 999.9 ppm
             1000 to 9999 ppm

Benzene – 0 to 200 ppm
Butadiene – 0 to 200 ppm

Resolution

VOCs – 0.05 ppm
              0.1 ppm
             1 ppm

Benzene – 0.05 ppm
Butadiene – 0.05 ppm

Response Time (T90)

< 3 s (for VOCs)

< 60 s (for Benzene and Butadiene)

Dimensions
(W x H x D)

10” x 3.0” x 2.5”
(25.5 x 7.6 x 6.4 cm)

Weight

26 oz (738g)

Temperature

-4° to 122°F (-20° to 50°C)

Humidity

0-95%  RH (non-condensing)

Battery

– Alkaline battery adapter

– Rechargeable external field-replaceable lithium-ion battery pack if specified

Operating Hours

16 hours of operation

Alarms

  • 95dB at 12” (30cm) buzzer and flashing red LED to indicate exceeded preset limits
  • High: 3 beeps and flashes per second
  • Low: 2 beeps and flashes per second
  • STEL & TWA: 1 beep and flash per second
  • Alarms latching with manual override or automatic reset
  • Additional diagnostics alarm and display message for low battery and pump stall

IP Rating

IP65 unit running

Calibration

Two-point  or three-point calibration for zero and span

Calibration memory for 8 calibration gases, alarm limits, span values and calibration dates

Sampling Pump

Internal, integrated flow rate at 400 cc/mn
Sample from 100’ (30m) horizontally and vertically

Low Flow Alarm

Auto shut-off pump at low-flow condition

Communication

Download data and upload instrument set-up from PC through charging cradle 
Wireless data transmission through built-in Bluetooth

Hazardous Area Approval

US and Canada: UL, cUL, Classified as Intrinsically Safe for use in Class I, Division 1 Groups A, B, C, D

Europe: ATEX Ex II 2G EEx ia IIC T4
IECEx: CSA 10.0005 Ex ia IIC/IIB GbT4

Warranty

3 years for 10.6 ev lamp, 1 year for 9.8 eV lamp, pump, battery, sensor and instrument

 